D24030 | Date 10/2014

Malfunctions, warning and error messages

6.3 Power failure

In case of a power failure, the appliance operates as follows:

In manual mode

After power supply has been restored, operation is continued with the parameters set. The

time and duration of the power failure are documented in the log memory.

In timer or programme mode

In timer or programme mode

In case of an interruption of the power supply of less than 60 minutes, the current pro-

gramme is continued from the point at which it was interrupted. For longer interruptions of

the power supply, all appliance functions (heating, fan etc.) are switched off.

In remote control mode:

The previous values are restored. If a programme has been initiated via remote control, it is

continued.