4 timer mode, 5 type of the slide-in unit (grid or shelf) – Memmert HPP, Peltier-cooled incubators IPP plus Constant climate chambers User Manual
Page 47

D24030 | Date 10/2014
47
Menu mode
7.3.4 Timer mode
Here, you can choose whether the digital backwards counter with
target time setting (see page 29) should run setpoint-de-
pendent or not. This determines whether the timer should not
start until a tolerance band of ±3 K around the set temperature
is reached (Fig. 25, B) or if it should start right after activation
(A).
t
Timer
t
Timer
A
B
°C/°F
t
6 K
Fig. 25
Timer Mode
A
Timer independent of setpoint: Timer starts right after activation
Fig. 25
Fig. 25
B
Timer setpoint-dependent: Timer does not start until tolerance band is reached
A
Timer independent of setpoint: Timer starts right after activation
A
Timer independent of setpoint: Timer starts right after activation
7.3.5 Type of the slide-in unit (Grid or Shelf)
Here, you have to set the type of the slide-in unit (grid or shelf)
used. The selection Shelf enables you to adjust the control
function to the different air flow characteristics in the interior
when using optional sliding shelves instead of the grids that are
part of the standard delivery.