Menu mode

7.3.4 Timer mode

Here, you can choose whether the digital backwards counter with

target time setting (see page 29) should run setpoint-de-

pendent or not. This determines whether the timer should not

start until a tolerance band of ±3 K around the set temperature

is reached (Fig. 25, B) or if it should start right after activation

(A).

t

Timer

t

Timer

A

B

°C/°F

t

6 K

Fig. 25

Timer Mode

A

Timer independent of setpoint: Timer starts right after activation

Fig. 25

Fig. 25

B

Timer setpoint-dependent: Timer does not start until tolerance band is reached

A

Timer independent of setpoint: Timer starts right after activation

A

Timer independent of setpoint: Timer starts right after activation

7.3.5 Type of the slide-in unit (Grid or Shelf)

Here, you have to set the type of the slide-in unit (grid or shelf)

used. The selection Shelf enables you to adjust the control

function to the different air flow characteristics in the interior

when using optional sliding shelves instead of the grids that are

part of the standard delivery.