Malfunctions, warning and error messages

Error description

Cause of error

Rectifying errors

See

Display T:E-3 in the

temperature display

TEMP

T:E-3

Set 37.0

°C

37

.4

°C

Temperature operating

sensor defective. The moni-

toring sensor takes over the

measurement function.

The appliance can

be temporarily

operated

Contact customer

service as soon as

possible

page 2

Error message AI E-3 in

the temperature display

TEMP

AI E-3

Set 37.0

°C

37

.4

°C

Temperature monitoring

sensor defective. The ope-

rating sensor takes over the

measurement function.

The appliance can

temporarily be kept

in service

Contact customer

service as soon as

possible

page 2

Error message E-3 in

the temperature display

TEMP

Set 45.0

°C

E-3

°C

Operating and monitoring

sensor defective

Switch off appli-

ance.

Remove the cham-

ber load

Contact customer

service

page 2

Error message E-6 in

the humidity display

HUMIDITY

Set 50

.0%rh

%rh

E-6

Humidity sensor defective

No humidity control

possible

Contact customer

service

page 2

When switching on

the appliance, the start

animation is displayed

in another colour than

white

Cyan

:

Not enough storage

space on the SD card

Red

:

The system files could

not be loaded

Orange

: The

fonts and images could

not be loaded

Contact customer

service

Contact customer

service

Download the firmware

update at memmert.

com and install it

page 2

page 2