Malfunctions, warning and error messages

6. Malfunctions, warning and error messages

Warning!

After removing covers, live parts may be exposed. You may receive

an electric shock if you touch these parts. Malfunctions requiring

work inside the appliance may only be rectifi ed by electricians. Ob-

serve the separate service manual for this.

Do not try to rectify appliance errors yourself but contact the MEMMERT customer service

department (see page 2) or an authorised service point.
In case of enquiries, please always specify the model and appliance number on the nameplate

(see page 13).

6.1 Warning messages of the monitoring function

If the acoustic alarm

IfIf

has been activated in the

Sound

menu (see

page 56) which is indicated by the speaker symbol

, the

alarm is additionally signalled by an intermittent acoustic signal.

If the confirmation key is pressed, the acoustic alarm can be

temporarily switched off until the next alarm event occurs.

6.1.1 Temperature monitoring

Description

Cause

Action

See

Temperature alarm and

"ASF" are displayed

TEMP

ASF Set 38.5

°C

40

.4

°C

Automatic tem-

perature monitor

(ASF) triggered

Check if the door is closed. Close

the door.
Extend the ASF tolerance band
If the alarm continues: Contact

customer service

page

36
page 2

Temperature alarm and

"TWW" are displayed

TEMP

TWW Set 38.5

°C

42

.4

°C

The adjustable

undertemperature

/ overtemperature

controller (TWW)

has assumed heat-

ing control.

Increase the difference between

the monitoring and setpoint

temperature – by either increas-

ing the max value of the temper-

ature monitoring or decreasing

the setpoint temperature.
If the alarm continues: Contact

customer service

page

35

page 2

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept