Operation and control
The status display shows you which operating mode or operating state the appliance is
currently in. The current operating state is highlighted in colour and indicated by the text
display:
Appliance
is in programme mode
■
Programme is stopped
Appliance is in manual operating mode
The example on the right shows the appliance in man-
ual mode, identified by the coloured hand symbol.
►
When the appliance is in timer mode,
Timer active
is
displayed:
►
When the appliance is in remote control mode, the
symbol appears in the temperature display:
5.4.4 Manual mode
In this operating mode, the appliance runs in permanent
operation at the values set on the ControlCOCKPIT.
Adjustment options
Adjustment options
As described in chapter 5.4.2 , you can set the following parameters after pressing the cor-
responding activation key (in any sequence):
Temperature
Temperature
Adjustment range: model dependent (see nameplate and technical
date on page 14)
Heating operation is indicated by the symbol.
symbol.
Cooling is indicated by the icon.
You can select °C or °F as the temperature unit displayed (see
page 46).
°C
22
.4
TEMP
°C
Set 37.0
The minimum temperature that can be reached depends on the surrounding conditions.
The appliances reach up to to 25 °C below room temperature. In order to do so, the
Peltier module needs sufficient ventilation (see Fig. 9 on page 19).
Humidity (for constant climate chambers HPP only)
Humidity (for constant climate chambers HPP only)
Adjustment range: 10 to 90 % rh
Humidifying is indicated by the symbol.
Dehumidifying is indicated by the icon.
HUMIDITY
70
%rh
Set
34
% rh
Interior lighting (only for models with light module)
Interior lighting (only for models with light module)
Adjustment range: 0 to 100 % in steps of 1%
LIGHT
%
40
Manual Mode
12.Sept.2012 13:44
Timer active
12.Sept.2012 13:44
23
.2°C
TEMP
Set 180
.0°C