7. Malfunctions: Meaning and Troubleshooting

Error No. /

INF No.

Error

Troubleshooting

Comment

InF 41

- Control board (C1/back

panel) and keyboard (C2)

have different software

versions.

- It is possible that due

to different software

versions, some or all

functions of the instru-

ment may be partly or

entirely unavailable. In-

form service and have the

software updated to the

latest version.

InF 42

- Control board (C1/back

panel) and VibroCheck

(C3) have different soft-

ware versions.

- It is possible that due to

different software ver-

sions, some or all func-

tions of the instrument

may be partly or entirely

unavailable.

"InF" messages can be removed by pressing the Clear key.

The message appears once each time the instrument is switched ON.

"InF" messages do not cause the instrument to be locked.

After all error messages, the instrument MUST be switched off, then on again using the main

switch. – Exceptions are InF 41 and 42.

Error 51

- Horizontal calibration of

VibroCheck is not pos-

sible.

- Possible operating error,

see the Instructions for

Use.

- Great deal of damage or

dirt on the blade or blade

holder. Use a new blade

or blade holder.

- Controllers have different

software versions. Have

the software updated.

- VibroCheck is defective.

Have it checked by ser-

vice.

- A calibration is carried

out even if the blade is

jagged.

A great deal of obvious

damage or dirt must be

present before a blade

calibration is canceled

automatically.