Instrument characteristics, 1 technical data for the vt1200, Technical data for the vt1200 – Leica Biosystems VT1200/VT1200 S User Manual
Page 10

8
Instructions for Use V 1.4 RevC – 09/2013
3.
Instrument characteristics
3.1 Technical data for the VT1200
General data:
Sectioning frequency (±10 %) ........................................................................................................... 85 Hz (± 10 %)
Amplitude.................................................................................................... from 0 - 3 mm, in 0.05 mm increments
Sectioning speed (± 10 %) ................................................................................................................ 0.01 - 1.5 mm/s
Return speed (± 10 %) ..................................................................................................................................2.5 mm/s
Total vertical specimen stroke .................................................................................................20 mm (motorized)
Cutting range ..............................................................................................................................45 mm (adjustable)
Maximum specimen size:
With standard blade holder .............................................................................................................33 x 50 mm
Specimen orientation, rotating ....................................................................................................................360°
Specimen plate, swiveling ....................................................................................................................... 0 - 10°
Section thickness adjustment .................................................................................. manual, in 1 µm increments
Ambient conditions:
Operating temperature range: ........................................................................................... min. 10 °C - max. 35 °C
Relative humidity: .......................................................................................................................................max. 60 %
Storage temperature: ...................................................................................................................................5 - 55 °C
Storage humidity: ............................................................................................................................................. < 60 %
Height: ........................................................................................................................ up to 2000 m above sea level
Electrical data:
Rated voltage range (± 10 %): ....................................................................................................................... 100 V - 240 V
Nominal frequency (±10 %): .................................................................................................................................. 50/60 Hz
Power consumption: ...................................................................................................................................................35 VA
Protective class: ...................................................................................................................................................................I
Power fuse: ...................................................................................................................................................... T 1 A L 250 V
Pollution degree: ..................................................................................................................................................................2
Overvoltage category: ....................................................................................................................................................... II
Electrical overload protection: ..................................................................................................................................... Yes
Internal current limit for the electronics: ................................................................................................................... Yes
Dimensions:
L x W x H: ............................................................................................................................... 600 mm x 250 mm x 230 mm
Height with magnifier support ............................................................................................ 600 mm x 250 mm x 320 mm
Height with microscope: ..................................................................................................... 600 mm x 250 mm x 490 mm
L x W x H control unit (when the bases are folded in): ................................................... 165 mm x 120 mm x 72 mm
Weight:
Without magnifier support and control unit .............................................................................................................56 kg
VT1200 control unit .........................................................................................................................................................1 kg
Magnifier support ...........................................................................................................................................................2 kg
Microscope support with stereomicroscope ............................................................................................................4 kg