beautypg.com

Instrument characteristics, 1 technical data for the vt1200, Technical data for the vt1200 – Leica Biosystems VT1200/VT1200 S User Manual

Page 10

background image

8

Instructions for Use V 1.4 RevC – 09/2013

3.

Instrument characteristics

3.1 Technical data for the VT1200

General data:

Sectioning frequency (±10 %) ........................................................................................................... 85 Hz (± 10 %)

Amplitude.................................................................................................... from 0 - 3 mm, in 0.05 mm increments

Sectioning speed (± 10 %) ................................................................................................................ 0.01 - 1.5 mm/s

Return speed (± 10 %) ..................................................................................................................................2.5 mm/s

Total vertical specimen stroke .................................................................................................20 mm (motorized)

Cutting range ..............................................................................................................................45 mm (adjustable)

Maximum specimen size:

With standard blade holder .............................................................................................................33 x 50 mm

Specimen orientation, rotating ....................................................................................................................360°

Specimen plate, swiveling ....................................................................................................................... 0 - 10°

Section thickness adjustment .................................................................................. manual, in 1 µm increments

Ambient conditions:

Operating temperature range: ........................................................................................... min. 10 °C - max. 35 °C

Relative humidity: .......................................................................................................................................max. 60 %

Storage temperature: ...................................................................................................................................5 - 55 °C

Storage humidity: ............................................................................................................................................. < 60 %

Height: ........................................................................................................................ up to 2000 m above sea level

Electrical data:

Rated voltage range (± 10 %): ....................................................................................................................... 100 V - 240 V

Nominal frequency (±10 %): .................................................................................................................................. 50/60 Hz

Power consumption: ...................................................................................................................................................35 VA

Protective class: ...................................................................................................................................................................I

Power fuse: ...................................................................................................................................................... T 1 A L 250 V

Pollution degree: ..................................................................................................................................................................2

Overvoltage category: ....................................................................................................................................................... II

Electrical overload protection: ..................................................................................................................................... Yes

Internal current limit for the electronics: ................................................................................................................... Yes

Dimensions:

L x W x H: ............................................................................................................................... 600 mm x 250 mm x 230 mm

Height with magnifier support ............................................................................................ 600 mm x 250 mm x 320 mm

Height with microscope: ..................................................................................................... 600 mm x 250 mm x 490 mm

L x W x H control unit (when the bases are folded in): ................................................... 165 mm x 120 mm x 72 mm

Weight:

Without magnifier support and control unit .............................................................................................................56 kg

VT1200 control unit .........................................................................................................................................................1 kg

Magnifier support ...........................................................................................................................................................2 kg

Microscope support with stereomicroscope ............................................................................................................4 kg