3.1.1 Technical data for the VT1200 S
General data:
Sectioning frequency (± 10 %): ...................................................................................................................85 Hz (± 10 %)
Amplitude:
.......................................................................................................from 0 - 3 mm, in increments of 0.05 mm
Sectioning speed (± 10 %): ........................................................................................................................0.01 - 1.5 mm/s
Return speed (± 10 %): ....................................................................................1.0 - 5 mm/s, in increments of 0.5 mm/s
Total vertical specimen stroke: ......................................................................................................... 20 mm (motorized)
Sectioning range: ..................................................................................................................................................... 45 mm
Sectioning window: .................................................................................................................................. 0.5 mm - 45 mm
Specimen retraction: ..............................................................................0 - 100 µm (adjustable; can be deactivated)
Maximum specimen size:
with standard blade holder: ..................................................................................................................... 33 x 50 mm
Specimen orientation, rotating: ...........................................................................................................................360°
Specimen plate, swiveling: ...............................................................................................................................0 - 10°
Sectioning thickness setting: ............................................manual in 1 µm increments or automatic max. 1000 µm
Ambient conditions:
Operating temperature range: .....................................................................................................min. 10 °C - max. 35 °C
Relative humidity: ............................................................................................................................................... max. 60 %
Storage temperature: .............................................................................................................................................5 - 55 °C
Storage humidity: .......................................................................................................................................................< 60 %
Height:
.................................................................................................................................up to 2000 m above sea level
Electrical data:
Rated voltage range (± 10 %): ...................................................................................................................... 100 V - 240 V
Nominal frequency (± 10 %): ................................................................................................................................ 50/60 Hz
Power consumption: ..................................................................................................................................................35 VA
Protective class: ..................................................................................................................................................................I
Power fuse: ..................................................................................................................................................... T 1 A L 250 V
Pollution degree: .................................................................................................................................................................2
Overvoltage category: ...................................................................................................................................................... II
Electrical overload protection: .................................................................................................................................... Yes
Internal current limit for the electronics: .................................................................................................................. Yes
Dimensions:
L x W x H: .............................................................................................................................. 600 mm x 250 mm x 230 mm
Height with magnifier: ......................................................................................................... 600 mm x 250 mm x 320 mm
Height with microscope: .................................................................................................... 600 mm x 250 mm x 490 mm
Control unit (when the bases are folded in): ..................................................................... 190 mm x 150 mm x 72 mm
Weight:
Without magnifier support and control unit: ...........................................................................................................56 kg
VT1200 S control unit: ...................................................................................................................................................1 kg
Magnifier support: .........................................................................................................................................................2 kg
Microscope support with stereomicroscope: ..........................................................................................................4 kg
3. Instrument Characteristics