Operation – Leica Biosystems IPC User Manual
Page 49

47
Leica IP C
Warning:
Problem with cassette ejection from a magazine
81 The display consists of two parts: The "8" is a warning that a magazine
ejector is mechanically blocked. The second digit of the message
specifies the number of the affected magazine.
Error messages
All displayed numbers between 20 and 78 and 89 and 93.
Display
14 Prompt requesting maintenance
If this message is displayed, the instrument will be due for main-
tenance within the next few weeks. Confirm the prompt pressing
ERROR.
After about 3 weeks the message will be displayed again and will not
disappear from the screen when pressing ERROR.
You can still continue printing but maintenance must be carried out
urgently.
15 Cleaning the print head
Screen prompt requesting the operator to manually clean the print
head (cleaning swab + alcohol).
The printer is off-line. No new print jobs are accepted.
87 After the last cartridge change, CLEAN
has been pressed to indicate to the printer that a cleaning cartridge
has been inserted. The printer has received a print job but is unable
to print becausethe cartridge contains cleaning fluid instead of ink.
Remedy:
Cancel the print job. Switch the printer off and back on and change
the cartridge. Then press LOADED or ERROR and wait for 2 minutes.
88 Ink cartridge has been changed, device waits for confirmation
with ERROR, CLEAN, or LOADED key.
Status messages (continued)
5. Operation
Caution!
NEVER press LOADED after reinserting an ink cartridge which has already
been used. This could cause permanent damage to the printer.