Leica IP C
Filling and inserting the magazines
(continued)
Magazine
positions
on the
Printer
6
3
2
1
5
4
Filling the magazines
• First fill IP C magazines 4 to 6 with cassettes
and insert them.
• Then fill IP C magazines 1 to 3 with cassettes
and insert them (
Fig. 31
).
Fig. 31
Fig. 32
The following table shows which cassette size
should be used with which inserts.
• When filling a magazine with cassettes, ensure
that the cassette side to be imprinted is to the
left after insertion.
• If cassettes are individually loaded, be certain
that they are aligned properly and that there
are no gaps between the cassettes.
• To achieve this, use your finger to carefully lift
the cassettes slightly from below, then release
them—they should then lie in the magazine
properly (
see Fig. 32
).
When filling the magazines, no
gaps may exist between the
cassettes.
4. Instrument Setup
Cassette size (length)
Insert
without lid or with
closed lid:
> 41.3 mm
Metal insert
without lid or with
closed lid:
< 39.3 mm
Metal strip and
adhesive strip at
front
with open lid:
< 80.0 mm Without any insert
with open lid:
< 77.3 mm Adhesive strip at
front