3 print specifications, 1 requirements for cassettes, Print specifications – Leica Biosystems IPC User Manual
Page 18: Instrument components and specifications

16
Instructions for Use V 1.9 RevD – 05/2013
3.3 Print specifications
Only standard histology cassettes that are reli-
ably inserted into the magazines can be used in
the Leica IP C.
3. Instrument Components and Specifications
Fig. 5
3.3.1 Requirements for cassettes
A wide variety of standard cassettes may be
used in the IP C; however, the following restric-
tions must be observed:
• Suitable for printing are all standard cassettes
without lids, of the following dimensions:
Length without lid x Width = max. 41.3 x
28.9 mm
Length with lid x Width = max. 80.0 x 28.9 mm
(Refer also to Technical Data,
p. 11
)
• Cassettes with lids attached have to be one-
piece units (23); the lids must not just be linked
to the body by a hinge.
• Cassettes with a flexible hinge cannot be used
unless the lid is detached (25) or closed.
• Cassettes with a closed lid (26) must be
checked to ensure that all four corners of the
lid are firmly closed and flat.
• Cassettes with top-mounted lids (24) cannot
be processed.
• For details on how to correctly insert the
cassettes into the magazines, please refer to
chapter 4.7
.
Fig. 7
22
22
23
24
25
26
26
Fig. 6