KD Scientific Allegro Peristaltic Pump Series User Manual
Page 73
73
K
D
S
A
lle
g
ro
P
e
ri
s
ta
lt
ic
P
u
m
p
S
y
s
te
m
U
s
e
r’s
M
a
n
u
a
l
phone
508.429.6809
fax
508.893.0160
web
www.kdscientific.com
KD Scientific Allegro Peristaltic Pump System User’s Manual
Publication 5613-004-REV-A
Appendices
TROUBLESHOOTING
1. Infusion Accuracy
To ensure infusion accuracy always use new tubing and
measure tubing bore diameter and enter actual
dimensions in millimeters (mm) using the Custom Tubing
entry option. Additionally, make sure that the guide rods
and lead screw are properly lubricated.
✐
NOTE: It is recommended that you save
your programs periodically by uploading
them to a PC.
KDS Allegro Peristaltic Pump System
Display Error Messages
2. Motor Stalled
This indicates that the pusher block travel has been
impeded. This may be caused by the tubing plunger
hitting bottom, a kink in the tubing (occlusion), tubing
plunger binding or any situation requiring more force to
the head of the tubing than the pump is capable of
delivering. To restart the pump, remove the cause of
stalling and/or increase the force setting, then press
RUN. The pump will continue the interrupted procedure
where it stopped.
3. Out of Range
A value was entered or encountered in a pump program
that was beyond the pump’s limits.
4. Power Failure Notification
If power to the pump is interrupted during operation, an
alarm will sound to alert the user to the interruption.
When power is restored, an onscreen message is
displayed to indicate the interruption.
5. Unit Won’t Power Up
Verify the the Power Source Switch on the rear of the
pump is set to DC-IN if powering through power supply
or RS-485 if powering in a chain.
The KDS Allegro Peristaltic Pump System Pump will notify you of a power
failure with a message when power is restored