Operating the KDS Allegro Peristaltic Pump System
K
D
S
A
lle
g
ro
P
e
ri
s
ta
lt
ic
P
u
m
p
S
y
s
te
m
U
s
e
r’s
M
a
n
u
a
l
phone
508.429.6809
fax
508.893.0160
web
www.kdscientific.com
KD Scientific Allego Peristaltic Pump System User’s Manual
Publication 5613-004-REV-A
I/O (IN) Event Link
The I/O (IN) Event Link command is used to have the
pump jump to a linked program if the signal is received.
The external signals are communicated via Pin 2 on the
15-pin I/O Connector.
TO INSERT AN I/O (IN) EVENT LINK COMMAND:
1.
From the Program Main screen, choose
Step Definition to enter the Step Definition screen.
2. Highlight (select once) the Program Step that you
want the I/O (IN) Event Link command to follow.
3. Choose the
button to access the
Program Steps Menu screen.
4. Choose
Advanced Options
to enter the Select
Advanced Options screen.
5. Choose the
I/O (IN) Event Link
button to enter the
On Event Step Definition Screen.
6. Select the desired Event type that you want to have
the event associated with from the options
available (Input Low, Input High, User Press, Motor
Stall, Withdraw, Infuse).
7.
After selecting the desired option in the On Event
screen, choose Accept to access the Program List
selection screen. Choose the Program name to
jump to. To use this command, you must have
defined a linkable Program.
8. Choose the Accept button to save your changes
and return to the Step Definition screen.
‘INS STEP’
The following options are available when setting an I/O
(IN) Event Link command. This command requires that
you have a program set with the specific actions (steps)
you want to perform if the event occurs so that you can
choose the Program name being linked by this command.
•
Input High-to-Low: Will link to the desired
program upon a logic low input signal falling edge.
•
Input Low-to-High: Will link to the desired
program upon a logic high input signal rising edge.
•
User Press: Will link to the desired program upon
a manual input.
• Motor Stall: Will link to the desired program
upon a Motor Stall event.
• Withdraw: Will link to the desired program upon
a Withdraw activity.
• Infuse: Will link to the desired program upon an
Infuse activity.
✐
NOTE: Since the ‘on event’ will cause a
jump (link) to a different program, be
cautious not to use embedded links in your
program programming.
Advanced Options I/O (IN) Event Link Selection Screen