7. Upon completion of the calibration run period,
the ‘Enter Volume Delivered’ button will be
activated. Please note the Volume reading listed
under the Tubing and Rate information on the
Calibration screen.
8. Select the Enter Volume Delivered button and
enter the Volume as measured on your scale
into the keypad and press Enter.
9. Press Enter on the Calibration screen to save
the setting or select the Run button to start the
procedure over again, or press the Reset
Calibration button to reset to the Factory
default value.