Nominal Minimum/Maximum Flow Rates for Various Syringes.
(Actual Limits will vary depending on syringe manufacturer)
Syringe Size
Syringe ID
Minimum Rate
Units
Maximum Rate
Units
0.5 µl
0.103 mm
3.66000
pl/min
1.90879
µl/min
1 µl
0.1457 mm
7.32000
pl/min
3.81946
µl/min
2 µl
0.206 mm
14.7000
pl/min
7.63515
µl/min
5 µl
0.343 mm
40.7400
pl/min
21.1676
µl/min
10 µl
0.485 mm
81.4800
pl/min
42.3220
µl/min
25 µl
0.729 mm
184.140
pl/min
95.6177
µl/min
50 µl
1.030 mm
367.560
pl/min
190.879
µl/min
100 µl
1.457 mm
735.600
pl/min
381.946
µl/min
250 µl
2.304 mm
1.83942
nl/min
955.098
µl/min
500 µl
3.256 mm
3.67356
nl/min
1.90744
ml/min
1 ml
4.608 mm
7.35774
nl/min
3.82039
ml/min