KD Scientific Legato 130 Series User Manual
Page 24

24
Configuring the KDS Legato 130
K
D
S
L
e
g
a
to
1
3
0
U
s
e
r’s
M
a
n
u
a
l
KD Scientific Legato 130 Syringe Pump User’s Manual
Publication 5617-008 Rev D
Setting the Force Limits
The maximum linear force of the syringe pump is 11 lbs
or 5 kgs. This force is sufficient to damage delicate, low
volume syringes, as well as many standard syringes if
pumping viscous fluids. The % of maximum force applied
by the pusher block can be changed.
In the set up screen, press
Force
.
Set the force level.
Press
Other Value
to enter a custom force %
Accept
the changes made by pressing
and go back to the Set Force screen.
Cancel
the changes made by pressing
and go back to the Set Force screen.
For a Custom Force %
Enter the precentage for custom force %
Accept
the changes made by pressing
and go back to the Set Force screen.
Cancel
the changes made by pressing
and go back to the Set Force screen.
✐
NOTE: A minimum value of approximately
30% is required to overcome friction forces
in the mechanism. You may have to set the
force level to a higher value depending on
the backpressure/friction in your system.
Press Here
Press here to enter
custom force %