14. Cleaning / repairing the unit
Caution:
Always turn off the unit and disconnect the mains cable from the power source
before cleaning the unit.
Prevent humidity from entering into the circulator.
Service and repair work may be performed only by authorized electricians.
Air cooled recirculating coolers:
To maintain the full cooling performance, clean the condenser
from time to time.
Switch off the unit, disconnect mains power cable.
Remove the venting grid.
Clean the ribbed condenser with a vacuum cleaner.
Replace the venting grid.
Switch on the unit.
Cleaning:
For cleaning the bath tank and the immersed parts of the recirculating
cooler, use low surface tension water (e.g., soap suds).
Clean the outside of the unit using a wet cloth and low surface tension
water.
The recirculating cooler is designed for continuous operation under
normal conditions. Periodic maintenance is not required.
The tank should be filled only with a bath fluid recommended by
JULABO. To avoid contamination, it is essential to change the bath fluid
from time to time.
Repairs
Before asking for a service technician or returning a JULABO
instrument for repair, please contact an authorized JULABO service
station.
When returning the unit:
Clean the unit in order to avoid any harm to the service personnel.
Attach a short fault description.
During transport the unit has to stand upright. Mark the packing
correspondingly.
When returning a unit, take care of careful and adequate packing.
JULABO is not responsible for damages that might occur from
insufficient packing.
JULABO reserves the right to carry out technical modifications with
repairs for providing improved performance of a unit.