Safety installations, warning functions

30

8.5. Change-over of the warning function to shutdown function


Factory setting: Li 0


t High

High temperature limit

t Low

Low temperature limit

For the two menu items > t High < and > t Low < choose between a
warning message being signalled or a complete shutdown of the main
functional elements such as heater and circulating pump being effected.
(see page 28)

 Setting >Li / 0<

An audible warning (interval tone) sounds and a meassage appears
on the MULTI-DISPLAY (LED).
t High

t Low

or

 Setting >Li / 1<

A complete shutdown of heater and circulating pump is effected.
An audible alarm (continuous tone) sounds and a message appears
on the MULTI-DISPLAY (LED).
t High

t Low

or

.

1.

Press the key

until the menu item

is displayed.

2.

Press enter

to indicate the adjusted parameter on the MULTI-

DISPLAY (LED) (digit blinks).

3.

Select the parameter with the keys

(0 / 1).

4.

Press enter

to store the selected parameter.

Press

to update the display immediately.