beautypg.com

JULABO SC 2500w Eco-Series Recirculating Coolers User Manual

Page 43

background image

Recirculating Cooler

43

in commands:

Asking for parameters or temperature values to be displayed.

Command

Parameter

Response of circulator

version

None

Number of software version (V X.xx)

status

none

Status message, error message (see page 44 )

in_pv_00

none

Actual bath temperature.

in_pv_01

none

Heating power being used (%).

in_pv_03

none

Temperature value registered by the safety sensor.

in_pv_04

none

Setpoint temperature („SafeTemp“) of the excess
temperature protection

in_sp_00

none

Working temperature „t 1“

in_sp_03

none

High temperature warning limit

„t High“

in_sp_04

none

Low temperature warning limit

„t Low“

in_sp_07

none

Pump pressure stage

in_par_00

none

Temperature difference between working sensor and safety
sensor

in_par_01

none

Te - Time constant of the external bath.

in_par_02

none

Si - Internal slope

in_par_06

none

Xp control parameter of the internal controller.

in_par_07

none

Tn control parameter of the internal controller.

in_par_08

none

Tv control parameter of the internal controller.

in_mode_05 none

Circulator in Stop/Start condition:

0 = Stop

1 = Start