beautypg.com

JULABO SC 2500w Eco-Series Recirculating Coolers User Manual

Page 12

background image

Operator responsibility – Safety instructions

12

Safety installations according to IEC 61010-2-010:
Excess temperature protection

adjustable from 0 °C ... 320 °C

Low liquid level protection

float switch

Classification according to DIN 12876-1

class III



Supplementary safety installations
Early warning system for low level

float switch

High temperature warning function

optical + audible (in intervals)

Low temperature warning function

optical + audible (in intervals)

Supervision of working sensor

plausibility control

Reciprocal sensor monitoring between
working and safety sensors

difference >35 K

Alarm message

optical + audible (permanent)

Warning message

optical + audible (in intervals)



Environmental conditions according to IEC 61 010-1:

Use indoors only.

Altitude up to 2000 m - normal zero.

Ambient temperature: see Technical specifications

Humidity:

Max. relative humidity 80% for temperatures up to +31 °C,

linear decrease down to 50% relative humidity at a temperature of +40 °C

Max. mains voltage fluctuations of ±10% are permissible.

Protection class according to IEC 60 529

IP21

The unit corresponds to Class I

Overvoltage category

II

Pollution degree

2

Caution:

The unit is not for use in explosive environment.


EMC requirements according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes), Class A
(industrial and commercial range).