JULABO SC 2500w Eco-Series Recirculating Coolers User Manual
Page 21
Recirculating Cooler
21
5.2.1. Tubing
Recommended tubing:
Maximum
pressure
Textile-reinforced tubing
> 4.5 bar
Warning:
Tubing:
At high working temperatures, the tubing used for temperature control and for the
cooling water supply represents a danger source.
A damaged tubing line may allow a large amount of hot bath fluid to be pumped out
within a short time.
This may result in:
Burning of skin
Breathing difficulties due to hot atmosphere
Safety instructions
Use suitable connecting tubing.
Make sure that the tubing is securely attached.
Avoid sharp bends in the tubing and maintain a sufficient distance from
surrounding walls.
Regularly check the tubing for material defects (e.g., for cracks), at least once a
year.
Preventive maintenance: replace the tubing from time to time.
5.2.2. Bath
fluids
Caution:
Carefully read the material safety data sheet of the bath fluid used, particularly with
regard to the fire point!
If a bath fluid with a fire point of
65 °C is used, only supervised operation is
possible.
Water:
The quality of water depends on local conditions.
Due to the high concentration of lime, hard water is not suitable for temperature
control because it leads to scale in the bath
Ferrous water can cause corrosion, even on stainless steel.
Chlorinated water can cause pitting corrosion.
Distilled water and deionized water are unsuitable. Their special properties
cause corrosion in the bath, even on stainless steel.