JULABO SC 2500w Eco-Series Recirculating Coolers User Manual
Page 42
Remote control
42
Important times for a command transmission:
To ensure a safe data transfer, the time gap between two commands should
be at least 250 ms.
The circulator automatically responds to an in command with a data string
followed by a LF (Line Feed). The next command should only be sent after
10 ms.
The out commands are valid only in remote control mode.
Examples:
Command to set the working temperature >Setpoint1< to 55.5 °C
out_sp_00
55.5
Command to ask for the working temperature >Setpoint1<
in_sp_00
Response from the circulator:
55.5
LF
12.3. List of commands
out commands:
Setting parameters or temperature values.
Command
Parameter
Response of circulator
out_mode_05
0
Stop the unit = R –OFF-.
out_mode_05
1
Start the unit.
out_sp_00
xxx.xx
Set working temperature. „t“
out_sp_03
xxx.xx
Set high temperature warning limit „t High“
out_sp_04
xxx.xx
Set low temperature warning limit „t Low“
out_sp_07
x
Set the pump pressure stage. (1 ... 3)
out_par_06
xxx
Xp control parameter of the internal controller.
out_par_07
xxx
Tn control parameter of the internal controller.
out_par_08
xxx
Tv control parameter of the internal controller.