Remote control

42

Important times for a command transmission:

To ensure a safe data transfer, the time gap between two commands should
be at least 250 ms.
The circulator automatically responds to an in command with a data string
followed by a LF (Line Feed). The next command should only be sent after
10 ms.

The out commands are valid only in remote control mode.


Examples:
Command to set the working temperature >Setpoint1< to 55.5 °C

out_sp_00

55.5

Command to ask for the working temperature >Setpoint1<

in_sp_00

Response from the circulator:

55.5

LF

12.3. List of commands

out commands:

Setting parameters or temperature values.

Command

Parameter

Response of circulator

out_mode_05

0

Stop the unit = R –OFF-.

out_mode_05

1

Start the unit.

out_sp_00

xxx.xx

Set working temperature. „t“

out_sp_03

xxx.xx

Set high temperature warning limit „t High“

out_sp_04

xxx.xx

Set low temperature warning limit „t Low“

out_sp_07

x

Set the pump pressure stage. (1 ... 3)

out_par_06

xxx

Xp control parameter of the internal controller.

out_par_07

xxx

Tn control parameter of the internal controller.

out_par_08

xxx

Tv control parameter of the internal controller.