High and low temperature warning functions – JULABO SC 2500w Eco-Series Recirculating Coolers User Manual
Page 28
Safety installations, warning functions
28
8.3. High and low temperature warning functions
Factory settings:
High temperature limit
t High
205 °C
Low temperature limit
t Low
-55 °C
If for a sensitive temperature application task adherence to a working
temperature value (setpoint) is to be supervised, then set high and low
temperature warning values.
In the example below, the >Setpoint< of 85 °C is surrounded by the
values > t High < 87 °C and > t Low < 83 °C. The electronics
immediately registers when the actual temperature attains a temperature
out of the limits and it follows a reaction according to what is set in the
menu item > Li < (Warning or Alarm).
(see chapter 8.5.
Change-over of the warning function to shutdown
function).
1.
Press the key
until the menu item
or
is displayed.
2.
Press enter
to indicate the adjusted value on the MULTI-
DISPLAY (LED) (digit blinks).
3.
Set value:
Use the cursor keys to move left or right on the display until
the numeral you wish to change blinks.
Use the increase/decrease arrows
to change the selected
numeral (-, 0, 1, 2, 3, ... 9).
4.
Press enter
to store the selected value.
Press
to update the display immediately
The warning functions are only triggered when the actual bath
temperature, after start from the „OFF“ or „rOFF“ mode, lies within
the set limits for 3 seconds.
Recommendation:
Set the high temperature warning value >t High< 5 °C to10 °C above the
working temperature setpoint
Set the low temperature warning value >t Low< 5 °C to 10 °C below the
working temperature setpoint