Technical specifications – JULABO SC 2500w Eco-Series Recirculating Coolers User Manual
Page 10
Operator responsibility – Safety instructions
10
2.4. Technical
specifications
Eco- Series
SC 2500a
SC 2500w
Working temperature ranges:
Standard unit
°C
+5 ... +35
+5 ... +35
Option Low Temp
°C
-20 ... +35
-20 ... +35
Option Low /HighTemp I
°C
-20 ... +80
-20 ... +80
Temperature stability
°C
0.1
0.1
A
bsolute Temperature Calibration ±3 ±3
Heater wattage (at 230 V) Option H1
kW
1.0
1.0
Cooling capacity
°C
Medium ethanol
kW
+20 0 -10
2.5 1.5 0.9
+20 0 -10
2.5 1.5 0.9
Refrigerant
R404A
R404A
Pump capacity P3
see table 1 page 11
Flow rate
P3
L/min at 0 bar
33
33
Pressure max.
P3
bar at 0 liters
3,5
3,5
Overall dimensions (WxDxH)
cm
49x62/105
49x62/105
Filling volume
liters
21 ... 33
21 ... 33
Weight
kg 135
122,5
Ambient temperature
°C
5 ... 40
5 ... 40
Mains power connection 195-253V/50Hz
V/ Hz
230 / 50
230 / 50
Current consumption
without heater
A
11 (230 V)
11 (230 V)
Current consumption
1kW heater
A
15 (230 V)
15 (230 V)
Mains power connection 195-253V/60Hz
V/ Hz
208-230 / 60
208-230 / 60
Current consumption
without heater
A
11 (230 V)
11 (230 V)
Current consumption
1kW heater
A
15 (230 V)
15 (230 V)
All measurements have been carried out at: rated voltage and frequency ambient temperature: 20 °C
Technical changes without prior notification reserved.
* Pump P3- reduces cooling capacity by 0.3 kW