2 technical data – INFICON HPG400 ATM to High-Vacuum Gauge User Manual
Page 8

8
tina31e1 (2004-05) HPG400 v1.om
2 Technical Data
Measuring range (air, N
2
)
Hot cathode
Pirani (control range)
2×10
-6
… 1 mbar
1×10
-2
… 1000 mbar
Accuracy
10
-5
… 1 mbar
±15% of reading
(valid between 10
-5
mbar and change-
over threshold setting (
↔ HP>,
→ 25).
Repeatability
10
-5
… 10
-1
mbar
10
-1
… 100 mbar
(after 10 min. stabilization)
±2% of reading
±30% of reading
Gas type dependence
→
Appendix B
Switching (changeover) threshold
(selectable from 5 defined setpoints)
1 mbar
5×10
–1
mbar
2×10
–1
mbar
1×10
–1
mbar
5×10
–2
mbar
Emission current
from 1 mbar
to 2×10
–6
mbar
continually rising
4 µA
130 µA
Anode voltage
180 V
Voltage range
0 … +10.2 VDC
Measuring range
Hot cathode
Pirani
1.5 … 7.5 V
8.5 … 9.75 V
Overrange hot cathode
Underrange hot cathode
Overrange Pirani
Underrange Pirani
7.5 V
≤ U ≤ 8.0 V
0.5 V
≤ U ≤ 1.5 V
9.75 V
≤ U ≤ 10.2 V
8.0 V
≤ U ≤ 8.5 V
Relationship voltage-pressure
Hot cathode
Pirani
logarithmic,
→
Appendix A
1.00 V/decade
0.25 V/decade
Error signals
Hot cathode error
Pirani error
≤+0.3 V, → 40
≤+0.5 V, → 40
Min. load impedance
10 k
Ω, short cicuit proof
Resistor
56 k
Ω between pin 10 and pin 5
(sensor cable connector)
Measurement
Emission
Output signal
Identification of the gauge
- TGF10 Tracer Gas Filler (36 pages)
- Sensistor ILS500 F Leak Detection Filler (90 pages)
- T-Guard Leak Detection Sensor (85 pages)
- T-Guard Leak Detection Sensor Interface description (40 pages)
- Sensistor ISH2000P Hydrogen Leak Detector, Panel Model (51 pages)
- Sensistor ISH2000 HySpeed Hydrogen Leak Detector (54 pages)
- LDS3000 Modular Leak Detector (52 pages)
- LDS3000 Modular Leak Detector Interface description (56 pages)
- BM1000 Bus module (14 pages)
- I/O1000 I/O module (18 pages)
- CU1000 Control unit (24 pages)
- Helium Leak Detector Modul1000 (130 pages)
- Helium Leak Detector Modul1000 Interface description (40 pages)
- UL5000 Dry Helium Leak Detector (108 pages)
- UL5000 Dry Helium Leak Detector Interface description (14 pages)
- UL1000 Fab Dry Helium Leak Detector (119 pages)
- HLD6000 Refrigerant Leak Detector (76 pages)
- HLD6000 Refrigerant Leak Detector Interface Description (40 pages)
- IO1000 I/O module (18 pages)
- Ecotec E3000 Multigas-Sniffer-Leak Detector (92 pages)
- Ecotec E3000 Multigas-Sniffer-Leak Detector Interface description (36 pages)
- Sensistor XRS9012 Hydrogen Leak Detector User Manual (28 pages)
- Sensistor XRS9012 Hydrogen Leak Detector Maintenance manual (14 pages)
- Extrima Ex-certified Hydrogen Leak Detector (62 pages)
- Sensistor ILS500 Leak Detection System (107 pages)
- Sensistor ISH2000 Hydrogen Leak Detector (108 pages)
- Sensistor ISH2000 Hydrogen Leak Detector (58 pages)
- Sensistor Sentrac Hydrogen Leak Detector (86 pages)
- Protec P3000(XL) Helium Leak Detector (132 pages)
- Pilot Plus Vacuum Gauge (2 pages)
- CO Check Carbon Monoxide Meter (2 pages)
- GAS-Mate Combustible Gas Leak Detector (12 pages)
- Whisper Ultrasonic Leak Detector (8 pages)
- Vortex AC Refrigerant Recovery Machine 115V (20 pages)
- Vortex AC Refrigerant Recovery Machine 230V (16 pages)
- Wey-TEK Refrigerant Charging Scale & Optional Charging Module (2 pages)
- Wey-TEK Refrigerant Charging Scale & Optional Charging Module (44 pages)
- D-TEK CO2 Refrigerant Leak Detector (12 pages)
- TEK-Mate Refrigerant Leak Detector (12 pages)
- Compass Refrigerant Leak Detector (12 pages)
- D-TEK Select Refrigerant Leak Detector (12 pages)
- Explorer Portable Gas Chromatograph (369 pages)
- MicroFID II Portable Flame Ionization Detector (89 pages)
- DataFID Portable Flame Ionization Detector for Landfill Emissions Monitoring (91 pages)
- Hydrostik Hydrogen Fuel Cylinder Installation (7 pages)