14

tira68e1 (2013-05) CDGxxxD.cp

Index

SI

DataType

NV Access PM Name

2800 0x01

BYTE RW Safe

State

0x02

REAL

RW

Safe

Value

0x03

BOOL

RW

Alarm

Enable

0x04

BOOL

RW

Warning

Enable

0x05 REAL

RW

Alarm Trip Point High

0x06 REAL

RW

Alarm Trip Point Low

0x07

REAL

RW

Alarm

Hysteresis

0x08 REAL

RW

Warning Trip Point High

0x09 REAL

RW

Warning Trip Point Low

0x0A

REAL

RW

Warning

Hysteresis

0x0B

BYTE

RW

Filter

Specifies the behavior for the value for States other than Valid.

Safe State

0
1
2
3

Zero
Full Scale
Hold last value
Use safe value

Safe Value: The value to be used for Safe State = Safe Value.

Alarm Enable: Enables the setting of the Alarm Status Bits.

Warning Enable: Enables the setting of the Warning Status Bits.

Alarm Trip Point High: Determines the value above which an Alarm Condition will
occur.

Alarm Trip Point Low: Determines the value below which an Alarm Condition will
occur.

Alarm Hysteresis: Determines the amount by which the value must recover to clear
an Alarm Condition.

Warning Trip Point High: Determines the value above which a Waning Condition
will occur.

Warning Trip Point Low: Determines the value below which a Waning Condition will
occur.

Warning Hysteresis: Determines the amount by which the value must recover to
clear a Warning Condition.

Filter

0
1
2

Dynamic (factory default)
Fast
Slow