beautypg.com

2 technical data – INFICON PCG55x ATM to Medium Vacuum Gauge User Manual

Page 10

background image

10

tina56e1-d (2014-01)

2 Technical

Data

For further technical data for gauges with DeviceNet and

Profibus interface

→  [3], [4], [5], [6].

Measurement range

5×10

-5

… 1500 mbar

Measurement principle

10 mbar … 1500 mbar


diaphragm capacitive
sensor

1 … 10 mbar

crossover range

5×10

-5

… 1 mbar

thermal conductance acc. to
Pirani

Accuracy (N

2

)

5Ч10

-4

… 1Ч10

-3

mbar


±50% of

reading

1×10

-3

… 100 mbar

±15% of reading

100 … 950 mbar

±5%

of reading

950 … 1050 mbar

±2.5% of reading

Repeatability (N

2

)

1×10

-3

… 1100 mbar


±2% of

reading

Output signal (measurement signal)

Voltage range
3PCx-0xx-xxx0


0 … +10.23 V

3PCx-0xx-xxx1

0 … +8.68 V

3PCx-0xx-xxx2

0 … +5.659 V

3PCx-0xx-xxx3

0 … +9.05 V

Measurement range
3PCx-0xx-xxx0


+0.61 … +10.23 V

3PCx-0xx-xxx1

+1.2 … +8.68 V

3PCx-0xx-xxx2

+0.375 … +5.659 V

3PCx-0xx-xxx3

+1.57 … +9.05 V

Error signal

0 V (default)