INFICON BPG402-SP ATM to Ultra-High Vacuum Gauge (Profibus) User Manual
Page 36

36
tira47e1 (2005-08) BPG402SPv1.cp
This service is used for activating degas and emission modes via Profibus.
ID
Name
Structure
Data type
Bytes Access
Range
14
Hot Cathode
Block State
Command
Rec
→ below
2
1/2_W
Byte 0
Transition
Command
Simple
UINT8
1
→ below
Byte 1
Value
Simple
UINT8
1
→ below
Transition
Command
Name
Description
0
Inactive
No action
1
Set Degas
State
Activates/deactivates the Degas mode
This service is used for activating/deactivating the
Degas mode (p <7.2
×10
-6
mbar).
If the Degas mode has not been deactivated with
Degas Off before, it is automatically turned off after
3 minutes.
The Degas State can only be changed if
p <7.2
×10
-6
mbar, otherwise the command returns the
error "Object State Conflict".
The Transition Command Data Field Value can have
the following values:
0
⇒ Degas Off
1
⇒ Degas On
3
Set
Emission
State
Turns the emission On or Off according to the target
value
0
⇒ Switch emission Off
1
⇒ Switch emission On
128
Emission
User Mode
The device supports two modes:
Automatic Mode:
Emission is switched on or off by the integrated Pirani
automatically.
If the emission is switched on and is then switched off
manually, the emission will be switched on again only
if the pressure rises above the 3.2
×10
-2
mbar and then
drops below 2.4
×10
-2
mbar.
Manual Mode:
Emission is switched on or off by the command
"Set Emission State"
• If during run time the pressure exceeds 3.2×10
-2
mbar, the emission is switched off automatically.
The measurement value equals the Pirani value.
• If the emission is switched off by the user while the
pressure is still in BA range, the measurement
value equals the Pirani value. There is no auto-
matic change to the BA range by the BPG402.
0
⇒ sets Manual Mode
1
⇒ sets Automatic Mode
3.3.3.12 Hot Cathode Block State
Command ID 14