3 cyclic message telegrams – INFICON BPG402-SP ATM to Ultra-High Vacuum Gauge (Profibus) User Manual
Page 11
tira47e1 (2005-08) BPG402SPv1.cp
11
The message telegrams listed below are implemented in the gauge. They can be
selected according to requirements. When selecting a message telegram, ascertain
what output format of the measured value (integer/float) is required and whether a
parameter channel is needed or not. The gauge can also be operated in such a
way that the master does not transmit any output data to the slave.
Standard
telegram
Master
⇔
Slave
Byte
Meaning
1
M
⇒ S
0
Transition Command
1
Transition Command Value for Hot Ion Gauges
2
M
⇒ S
0 … 7
Parameter Channel
8
Transition Command
9
Transition Command Value for Hot Ion Gauges
3
M
⇒ S
0 … 7
Parameter Channel
4
S
⇒ M
0
Exception status
1
One Of N status extension
2
One Of N PV selector
3 … 4
Process value UINT16
5
S
⇒ M
0
Exception status
1
One Of N status extension
2
One Of N PV selector
3 … 6
Process value float
6
S
⇒ M
0 … 7
Parameter channel
8
Exception status
9
One Of N status extension
10
One Of N PV selector
11 … 12 Process value UINT16
7
S
⇒ M
0 … 7
Parameter channel
8
Exception status
9
One Of N status extension
10
One Of N PV selector
11 … 14 Process value float
Depending on the standard telegrams used, the respective configuration data have
to be transmitted to the gauge during the configuration phase (
→ table on 5).
2.3 Cyclic Message
Telegrams
Configuration data