INFICON BPG402-SP ATM to Ultra-High Vacuum Gauge (Profibus) User Manual
Page 27
tira47e1 (2005-08) BPG402SPv1.cp
27
Instance 2 of the Analog Sensor Input Function Block describes the functionality of
the BA measuring part of the gauge.
In Instance 2, the same attributes as in Instance 1 are supported:
ID
Name
Structure Data type
Bytes
Access
Store
19
Process Value
(PV)
Simple
According to
Data Type value
(parameter 21)
-
0_XI
1/2_R
V
20
Status
Simple
UINT8
1
0_XI
1/2_R
V
21
Data Type
Simple
UINT8
1
2_R/W
N
22
Data Units
Simple
UINT16
2
2_R/W
N
23
Reading Valid
Simple
Boolean
1
1/2_R
V
24
Full Scale
Simple
According to
Data Type value
(parameter 21)
-
1/2_R
N
39
Safe State
Simple
UINT8
1
1/2_R/W
N
40
Safe Value
Simple
According to
Data Type value
(parameter 21)
-
1/2_R/W
N
44
Overrange
Simple
According to
Data Type value
(parameter 21)
-
1/2_R
N
45
Underrange
Simple
According to
Data Type value
(parameter 21)
-
1/2_R
N
The Process Value contains the measured value of the BA Device Instance in the
currently selected data unit (ID 22) and in the selected data type (ID 21).
If the device is not in the EXECUTING status (ID 25, Device Block), the value
defined in the Safe State is output.
Values output in the data unit COUNTS can be converted into a pressure value by
means of the following formulas:
Calculation of the pressure (PV is the abbreviation used for Process Value):
PV
mbar
=
10
(COUNTS / 2000) - 12.5
PV
Torr
=
0.75006168 × PV
mbar
PV
Micron
=
10
-3
× PV
Torr
PV
Pa
=
100 × PV
mbar
This parameter is remains on "0".
Two data types are supported: Float and Integer16.
In cyclic data traffic, the data type cannot be modified. After the gauge is switched
on, the value stored in the EEPROM is loaded. The data type setting can only be
modified when the gauge is in the IDLE status. By defining the configuration data
for cyclic data traffic (selection of standard telegram), the data type used in the
selected standard telegram is taken over. All settings previously made in acyclic
data traffic are thus overwritten.
3.2.3 Analog Sensor Input
Function Block
SLOT 1 / Instance 2
3.2.3.1 Process Value ID 19
3.2.3.2 Status ID 20
3.2.3.3 Data Type ID 21