INFICON BPG402-SD ATM to Ultra-High Vacuum Gauge (DeviceNet) User Manual
Page 6
6
tira46e1 (2004-11) BPG402-SDv1.cp
8
9
10
Position
Function
8
Potentiometer for Setpoint A threshold adjustment
9
Potentiometer for Setpoint B threshold adjustment
10
Fullscale adjustment push button
Device State
LED State
Description
Power Off
dark
No power applied to device.
Device Self-Test
flashing
green-Red
Device is in self-test.
Device Operational
green
Device is operating normally.
Unrecoverable
Fault
red
Device has detected an unrecoverable fault.
As stated previously, all module level faults
are considered as unrecoverable faults.
Minor Fault
flashing red
Device has detected a recoverable fault
(e.g. DeviceNet power supply not
connected).
The network status LED indicates the status of the DeviceNet communication link.
For this state
LED state
To indicate
Not Powered/Not Online
dark
Device is not online.
• The device has not completed the
Duplicate MAC_ID test yet.
• The device may not be powered,
look at Module Status LED.
On–line, Not Connected
flashing
green
Device is online but has no connections
in the established state.
• The device has passed the
Dup_MAC_ID test, is online, but
has no established connections to
other nodes.
• The device is not allocated to a
master.
Link OK Online, Connected
green
The device is online and has connec-
tions in the established state.
• The device is allocated to a Master.
Connection Time Out
flashing
red
One or more I/O Connections are in the
timed out state.
Critical Link Failure
red
Failed communication device. The
device has detected an error that has
rendered it incapable of communicating
on the network (Duplicate MAC ID, or
Bus–off).
1.4 Side View of the
BPG402-SD
1.5 Indicators and Switches
1.5.1 Module Status LED
1.5.2 Network Status LED