4 s-analog sensor object, 5 s-device supervisor object specific services, 1 class attributes – INFICON BPG402-SD ATM to Ultra-High Vacuum Gauge (DeviceNet) User Manual
Page 16

16
tira46e1 (2004-11) BPG402-SDv1.cp
Service
Code
Service Name
Description of Service
4B
h
Abort
Moves the device to the Abort state
4C
h
Recover
Moves the device out of the Abort state
4D
h
Perform_Diagnostics
Causes the device to perform a set of diag-
nostic routines
• DS Object Service Parameter dictionary
Parameter
Form
Description
TestID
USINT
Type and possibly detail of diagnostic test to be
performed
• TestID parameter
The following values are defined for the TestID parameter for the
Perform_Diagnostics Service Request:
Attribute Value
State
0
Standard
Type "Standard" is specified if there is only one type of diagnostic defined or if
there are more than one including a type standard.
Class Code 31
h
= 49
d
Attribute ID
Access
Rule
Name
DeviceNet
Data Type
Description of Attribute
Semantics of Values
1
Get
Revision
UINT
Revision of this object Note: All class definitions
are required to include this class attribute.
The current value
assigned to this attrib-
ute is one (01).
2
Get
Max
Instance
UINT
Maximum instance number of an object currently
created in this class level of the device.
The largest Instance
number of a created
object at this class
hierarchy level.
32
1)
Get
Class
Level
Status
Extension
USINT
Indicates whether all instances measure above or
below its maximum or minimum measurement
range
Bit 0: Reading Invalid
Bit 1: Overrange
Exceeded
Bit 2: Underrange
94
Get
Active
Value
Specified
by Data
Type
Is used by assemblies to produce this class-level
attribute, instead of the Value (Attribute ID 6) of the
S-Analog Sensor Instances.
95
Get
Active
Instance
Number
UINT
Identifies the object instance that is providing the
Value which is copied into the Active Value for all
input Assemblies and the Alarm/Warning Exception
Details for the S-Device Supervisor object.
→ Behavior section.
Default = 1
96
Get
Number of
Gauges
USINT
Identifies the number of gauge instances present in
the device.
2
99
Get
Subclass
UINT
Identifies a subset of additional class attributes,
services and behaviors.
1
Instance Selector
1)
This attribute is not part of the actual S-Analog Sensor Specificiation, but is at
this time in preparation to be implemented as shown. Changes may be possible
until the new specification is published.
2.3.5 S-Device Supervisor
Object Specific Services
Explanation to Service code
4D
h
2.4 S-Analog Sensor Object
2.4.1 Class Attributes