beautypg.com

INFICON BPG402-SD ATM to Ultra-High Vacuum Gauge (DeviceNet) User Manual

Page 21

background image

tira46e1 (2004-11) BPG402-SDv1.cp

21

Attribute

ID

Access

Rule

NV/
V

Name

DeviceNet
Data Type

Description of Attribute

Semantics of Values

96

Get

V

Status
Extension

BYTE

Bit mapped byte providing
additional status bits

Bit description:

0

Reading Invalid
(Logical Inversion of
Reading Valid)

1

Overrange Exceeded

2

Underrange Exceeded

99

Get

NV

Subclass

UINT

Identifies the subset of
additional attributes, serv-
ices and behaviors for hot
cathode ion gauges

5

Hot Cathode Ion Gauge

100

Get

V

State
Emission
User
Mode

USINT

Represents the state ac-
cording to the service Set
Emission User Mode

0

Auto

1

Manual





→ Instance 1

An S-Analog Sensor object instance derives a reading from a physical analog sen-
sor. The reading is converted to the data type and units specified for the Value
attribute.

Using Counts and INT the following conversion has to be used:

Counts

=

[ log

10

(pressure) + k ] × 2000

where:

k

mbar

=

12.5

k

Pa

=

10.5

k

Torr

=

12.624903



→ Instance 1

→ Instance 1

16 Bits are used as sensor faults . Bit 8

… Bit 16 are mapped to the Exception

Detail Alarm 3, Bit 0

… Bit 7 are mapped to the Exception Detail Alarm 2.

Data Component

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Sensor Alarm Byte 0

0

0

0

0

0

0

Sensor

Filament 2

Failure

Sensor

Filament 1

Failure

Sensor Alarm Byte 1

0

0

0

0

0

0

Electronics

Failure

0


2.4.2.4 Semantics of S-Analog

Sensor Instance 2

Data Type

Value

Safe State

Safe Value

Sensor Alarm