44

tinb07e1-e 2011-07 Vacuum Gauge Controller

6.3.36 TLC

Torr lock. See Chapter 5.4.4 Torr lock (tr-L), 30.

S: TLC[,a][]

R:

S:

R: a

6.3.37 TRA

Test RAM. Test the RAM. See Chapter 5.4.6 RAM test
(rA-t), 30
.

S: TRA[]

R:

S:

R: aaaa

The control character starts the test. It takes
approximately one second to complete the test.

6.3.38 TRS

Test RS232C interface. Test the RS232C interface. See
Chapter 5.4.13 RS232C test (rS-t), 31.

S: TRS[]

R:

S:

The control character starts the test. The test can
be stopped by pressing Ctrl-C.

6.3.39 UNI

Unit of measurement. See Chapter 5.3.1 Unit of mea-
surement (unit), 27
.

S: UNI[,a][]

R:

S:

R: a

6.3.40 WDT

Watchdog control. See Chapter 5.4.3 Watchdog control
(dt-C), 29.

S: WDT[,a][]

R:

S:

R: a

Parameter

Significance

a

Torr lock

0 = Off (default)
1 = On

Parameter

Significance

aaaa

Error status. See Chapter 6.3.9
ERR, 36
.

Parameter

Significance

a

Unit of measurement

0 = mbar/bar (default)
1 = Torr
2 = Pascal
3 = Micron

Parameter

Significance

a

Error acknowledgement

0 = Manually
1 = Automatic (default)