22 pre, 23 prx, 24 res – INFICON VGC403 Three-Channel Measurement and Control Unit User Manual
Page 40: 25 sav

40
tinb07e1-e 2011-07 Vacuum Gauge Controller
6.3.22 PRE
Pirani range extension. See Chapter 5.2.12 Pirani range 
extension (PrE),  26.
S: PRE[,a,b,c]
R: 
S: 
R: a,b,c
6.3.23 PRX
Pressure sensors. Pressure readings of all sensors.
S: PRX
R: 
S: 
R: a,±b.bbbbE±bb,c,±d.ddddE±dd,e,±f.ffffEff
6.3.24 RES
Reset. Reset the serial interface.
Deletes the input buffer. All queued error messages are 
sent to the Host. The unit returns to the measurement 
mode.
S: RES[,a]
R: 
S: 
R: b,c,d,...
6.3.25 SAV
The command SAV,0 resets all parameters to their 
default values (factory settings). See Chapter 5.3.4 
Default parameters (dEF),  27.
The command SAV,1 saves parameter values which 
have been changed via the serial interface in the 
EEPROM. These values will be preserved even when the 
unit is switched off.
S: SAV,a
R: 
Parameter
Significance
a
Range extension for 
Sensor 1
0 = Off (default)
1 = On
b
Range extension for 
sensor 2 (see above)
c
Range extension for 
sensor 3 (see above)
Parameter
Significance
a
Status of channel 1
0 = Measurement data ok
1 = Underrange
2 = Overrange
3 = Sensor error
4 = Sensor switched off
5 = No sensor
6 = Identification error
7 = BPG/BCG/HPG error
±b.bbbbE±bb
Pressure reading of sensor 1 in the 
current unit of measurement
c
Status of channel 2
(see above)
±d.ddddE±dd
Pressure reading of sensor 2 in the 
current unit of measurement
e
Status of channel 3
(see above)
±f.ffffE±ff
Pressure reading of sensor 3 in the 
current unit of measurement
Parameter
Significance
a
1 = Perform reset
b,c,d...
Queued error messages
0 = No error
1 = Watchdog control has been trig-
gered
2 = Task(s) not executed
3 = EPROM error
4 = RAM error
5 = EEPROM error
6 = Display error
7 = A/D converter error
8 = UART error
9 = Sensor 1 general error
10 = Sensor 1 ID error
11 = Sensor 2 general error
12 = Sensor 2 ID error
13 = Sensor 3 general error
14 = Sensor 3 ID error
Parameter
Significance