5.4.11 A/D converter ID (Ad-i)
For each channel, display a signal (in volts) which is 
caused by a resistor inside the connected sensor. This 
signal is used for identification of the connected sensors.
Press the UP button to start the test.
5.4.12 I/O test (io-t)
Test all relays. In this test the relays are switched on and 
off one after the other, and the relay states are shown on 
the display. A circuit indicator or an ohmmeter may be 
used to verify the relay states on the RELAY connection. 
See Chapter 3.3.5 RELAY,  13.
Press the UP button to start the test.
Tab. 5-18 Relay test
5.4.13 RS232C test (rS-t)
Test the RS232C interface. In this test, the Vacuum 
Gauge Controller echoes each character received from 
from the serial interface back to the interface.
The test starts as soon as the rS-t parameter has been 
selected. You may e.g. use a terminal program and input 
characters via the keyboard. Each of the input characters 
is returned to the terminal from the Vacuum Gauge Con-
troller. The data transfer between the two units is visible 
on the terminal screen only.
Press the PARA button to quit the test and to return to the 
measurement mode.
CAUTION
Relay test.
In this test the relays switch irrespective of the 
actual pressure. This may cause unintended 
switching of devices.
Unplug the RELAY connection before per-
forming a relay test.
Display
Significance
off
All relays switched off
r1-H
r1-L
Switching function 1 relay on
Switching function 1 relay off
r2-H
r2-L
Switching function 2 relay on
Switching function 2 relay off
r3-H
r3-L
Switching function 3 relay on
Switching function 3 relay off
r4-H
r4-L
Switching function 4 relay on
Switching function 4 relay off
r5-H
r5-L
Switching function 5 relay on
Switching function 5 relay off
r6-H
r6-L
Switching function 6 relay on
Switching function 6 relay off
r7-H
r7-L
Error signal relay on
Error signal relay off