2 interfacing to ic6 deposition controller – INFICON CI-100 Crucible Indexer User Manual
Page 25

2 - 9
IP
N 07
4-
54
4-
P1
B
CI-100 Crucible Indexer Operating Manual
On the CI-100 rear panel, set Select Switch #5 Down. On the SQC-310 System
Menu / Sensors & Sources, set up the Source with:
Number of Positions: 4
Control Type: Individual
Feedback Type: In Position
Indexer Delay: 5 seconds
2.8.2 Interfacing to IC6 Deposition Controller
The INFICON IC6 does not support binary indexer wiring. BCD wiring between the
CI-100 and the IC6 is required.
The cable wiring below interfaces the IC6 I/O Board connectors to the CI-100 BCD
I/O connector:
IC6 Relay
CI-100 BCD I/O
Pin 6 (Relay 1, NO)----->----Pin 1
Pocket Bit1 Select
Pin 8 (Relay 2, NO)----->----Pin 2
Pocket Bit2 Select
Pin 10 (Relay 3, NO)---->----Pin 7
Pocket Bit3 Select
Pin 7,9,11(Relay Common)-Pin 6
Common
IC6 Input
CI-100 BCD I/O
Pin 15 (Input 1)------<--------Pin 5
Pocket Ready B
Pin 1---------------------------- Pin 6
Common
Short Pin 3 to 9
Interlock
Short Pin 4 to 8
Pocket Ready A
On the CI-100 rear panel, set Select On/Off Switch #5 UP and #7 DOWN. Turn the
CI-100 Off, then On, to read the switch change.
On the IC6 Source menu select Number of Crucibles=8, Crucible Output = 1, and
Crucible Output Type =NO. Set Turret Feedback = Yes and Turret Input = 1.