Profibus-DP Protocol
5.1
Preface
In order to use Profibus-DP with T-Guard, you must have ordered a T-Guard in the
Profibus version (Cat. No. 540-002). It is not possible to add the Profibus option to
Cat. No. 540-001.
When setting up the Profibus communication you need to use the GSD file provided
by INFICON.
5.2
Process Data Mapping for Cyclic Data Transfer
5.2.1
Read Process Data (PLC-> Leak Detector)
This data is sent periodically from the programmable logic controller to the leak
detector:
Notice
The Profibus-DP protocol is subject to change.
If you are using this protocol, please ask INFICON for an update.
Byte Bit
Name
Meaning
Similar to PLC
Input
Similar to RS232
ASCII cmd.
0
not used
1
not used
2
CLEAR
Transition 0->1:Clears errors and
warnings
CLEAR
*CLS
3
not used
4
CAL
Transition 0->1: Start calibration
CAL
*CAL:START
5
PROOF
Transition 0->1: Start proof
PROOF
6
STOP
Transition 0->1: Stop
STOP
*STOP
7
START
Transition 0->1: Start
START
*START
0
PURGE_STOP
Transition 0->1 = Stop purging the
measurement line
*PURGE:STOP
1 PURGE_START
Transition 0->1 = Start purging the
measurement line
PURGE
*PURGE:START
2
not used
3
not used
4
STANDBY
Transition 0->1: Switch to
stand-by
STANDBY
*STANDBY
5
GROSS/FINE
Switch between GROSS and
FINE (in Cont. Mode only)
GROSS/FINE
*GROSS ON
6
READY
Transition 0->1: Wake up leak
detector from stand-by mode
READY
*STANDBY:STOP
7
not used