Manual set point controller (cont.) – Despatch 1500 Controller User Manual

Page 13

background image

Despatch/Watlow 1500 Control Instruction Manual

Page 7

NOTE:
In the DATA display the right
most character for Event 11
is shown as 0001.

Manual Set Point Controller (Cont.)

4.

Enter the %RH. The humidity event must be off when
outside of the range 4

E

C to 98

E

C. Humidity beyond

these extremes can result in equipment damage. If a
dry air system and an electronic humidity sensor are
installed in lieu of a wet bulb sensor, the optional dry
air event may be outside the range of 4

E

C to 98

E

C.

a. Press the FCTN key. C2 is shown.

b. Update the %RH temperature by using the up/down

keys.

c. Press enter when DATA display shows your desired

set point.

5.

Change the auxiliary events.

a. Press the FCTN key. E1 is displayed.

b. Turn the auxiliary event on or off.

1 = on
0 = off

Try varying set points and events and watch how each
effects the output LEDs.