Manual set point controller (cont.) – Despatch 1500 Controller User Manual
Page 13
Despatch/Watlow 1500 Control Instruction Manual
Page 7
NOTE:
In the DATA display the right
most character for Event 11
is shown as 0001.
Manual Set Point Controller (Cont.)
4.
Enter the %RH. The humidity event must be off when
outside of the range 4
E
C to 98
E
C. Humidity beyond
these extremes can result in equipment damage. If a
dry air system and an electronic humidity sensor are
installed in lieu of a wet bulb sensor, the optional dry
air event may be outside the range of 4
E
C to 98
E
C.
a. Press the FCTN key. C2 is shown.
b. Update the %RH temperature by using the up/down
keys.
c. Press enter when DATA display shows your desired
set point.
5.
Change the auxiliary events.
a. Press the FCTN key. E1 is displayed.
b. Turn the auxiliary event on or off.
1 = on
0 = off
Try varying set points and events and watch how each
effects the output LEDs.