beautypg.com

Dakota Ultrasonics MMX-7 User Manual

Page 9

background image

MMX-7 Data Logging Ultrasonic Thickness Gauge

5

2) Use the UP and DOWN arrow keys to scroll through the sub menu items

until VELOCITY is highlighted.

3) Press the ENTER key to display the Digits Edit Box.

4) Press the UP and DOWN arrow keys to scroll the highlighted value.

5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.

6) Repeat steps 4 & 5 until the velocity number is correctly displayed.

7) Press the OK key to set the velocity and return to the menu screen, or ESC

to cancel entering the velocity.

8) Finally, press the MEAS key to return to the measurement screen and begin

taking readings.