12) Press the OK key to activate the confirmation screen.

13) Press the OK key to save the Setup, or ESC to cancel saving the Setup.

14) Finally, press the MEAS key to return to the measurement screen.

Note: The Name and Comments of the Setup can be edited at any time by

simply repeating the Save Setup routine described above. Therefore, the

Save Setup function can also be considered an Edit Function.

11.4 Using the Default Setup

The default setup feature was added to the MMX-7 to use, as a last resort, if there
are no setups stored in the gauge –factory or otherwise. The only time this might
possibly occur is if the setup file in the MMX-7 was somehow corrupted, and the user
does not have access to a computer to re-load the factory setups back into the MMX-
7
. This gives the user the ability to load and modify a basic setup as follows:

Using the Default Setup