Control Company 6404 MEMORY CARD THERMOMETER/HYGROMETER/DEW POINT METER User Manual
Control Company Measuring instruments
SPECIFICATIONS
Temperature—
Range:
32.0 to 122.0 °F / 0.0 to 50.0 °C
Resolution:
0.1°
Humidity—
Range:
0.0 to 99.0% RH
Resolution:
0.1%
Dew Point—
Range:
-4.0 to 158.0 °F / -20.0 to 70.0 °C
Resolution:
0.1°
Display Update: 10 seconds
Memory Card: Up to 2GB SD / Up to 4GB SDHC
Data Logging: 1 minute to 720 minute user
selectable interval
LED visual alarm indicator
Ambient Temperature (TA)
Battery status indicator
Ambient Relative Humidity
Dew Point Temperature
Time-of-day clock (data logger messages will
also alternate here)
AC POWER SUPPLY/BATTERY POWER
The meter may be powered the following ways:
Batteries only
Typical operating time when using the three (3) each
AA alkaline batteries only:
Up to 1 year in display mode only, no data logging,
no alarms.
Up to 1 month with data logging on at a 1 minute
recording rate, with no alarms. Higher capacity
memory cards require more power to operate;
therefore, larger cards will exhaust the batteries
more rapidly.
AC power supply only
If the AC power supply is being used without
batteries, in the event of a power outage, the meter
will turn off and data logging will stop.
AC power supply and batteries together
Using the AC power supply and batteries together
allows the batteries to act as a backup power supply
in the event of a power outage.
When using the AC power supply and batteries
together, the batteries will need to be replaced
periodically to insure they have suf cient power to
operate the meter during a power outage (see the
“Battery Replacement” section).
DISPLAY MESSAGES
The following messages may appear alternating in
place of the time-of-day clock:
Sd
Indicates that an SD memory card is
inserted; however, data logging is
turned off.
rEc
Indicates that an SD memory card is
inserted and data logging is turned on.
FuLL
Indicates that the SD memory card is full
and data logging if off (no records may
be written to a full SD memory card)
Err & Sd Indicates that the SD memory card was
removed while data logging was turned on,
the meter is no longer recording data.
SETTING TIME-OF-DAY/DATE
1. Turn the meter on by pressing the ON/OFF
button.
2. Press and hold the SET and buttons
simultaneously for 3 seconds to enter the
time-of-day/date setting mode (12H or 24H will
ash on the display).
3. Press the or button to select 12 or 24 hour
time format.
4. Press the SET button, the year will ash.
5. Press the or button to increment the
display to the correct value. This is the last 2
digits of the year.
Example: 2013 is displayed as 13.
6. Press the SET button, the month will ash.
7. Press the or button to increment the
display to the correct value.
8. Press the SET button, the day number will ash.
9. Press the or button to increment the
display to the correct value.
10. Press the SET button, the time-of-day hours will
ash.
11. Press the or button to increment the
display to the correct value.
12. Press the SET button, the time-of-day minutes
will ash.
13. Press the or button to increment the
display to the correct value.
14. Press the SET button, the display will return to
the current temperature/relative humidity display.
DISPLAYING °F OR °C
With the meter turned on, to toggle the display
between reading Fahrenheit or Celsius, press and
hold the button for 3 seconds.
When using the data logging function, the
temperature values recorded will match the selected
display unit (°F or °C).
MINIMUM AND MAXIMUM MEMORY
There are six points that are automatically recorded
into memory:
• Minimum Ambient Temperature Achieved
• Maximum Ambient Temperature Achieved
• Minimum Ambient Relative Humidity Achieved
• Maximum Ambient Relative Humidity Achieved
• Minimum Dew Point Temperature Achieved
• Maximum Dew Point Temperature Achieved
Minimum and maximum memories are NOT
programmable. The minimum value recorded into
memory is the minimum reading achieved since the
last time the memory was cleared. The maximum
value recorded into memory is the maximum reading
achieved since the last time the memory was
cleared. The minimum and maximum memories are
maintained over the period since the memory was
last cleared.
VIEWING MINIMUM/MAXIMUM MEMORY
Press the MIN/MAX button, the display will show
“MIN” at the bottom left side of the display indicating
that the minimum memory values are being
displayed.
Press the MIN/MAX button a second time, the
display will show “MAX” at the bottom left side of the
display indicating that the maximum memory values
are being displayed.
Press the MIN/MAX button a third time to return to
the current temperature/humidity display (MIN/MAX
will no longer appear on the display).
CLEARING THE MINIMUM/MAXIMUM MEMORIES
Clearing the Minimum/Maximum memories will
set both the minimum and maximum values for all
readings to the current display readings.
The meter does not have to be in the Min/Max
memory display mode to clear the minimum/
maximum memories.
To clear the minimum and maximum memories,
press and hold the MIN/MAX button for 3 seconds,
all of the LCD segments will appear on the display for
approximately 3 seconds indicating that the minimum
and maximum memories have been cleared.
The minimum and maximum memories are also
cleared when the meter is turned off.
ALARMS
Independent High/Low alarms may be set for all
readings (ambient temperature, relative humidity and
dew point temperature). Temperature alarm limits
may be set in 0.1° increments. Relative humidity
alarm limits may be set in 0.1% RH increments.
Each High/Low alarm limit may be independently
enabled/disabled allowing selecting of high alarm
only, low alarm only, or both high and low alarm
monitoring.
SETTING HIGH/LOW ALARM LIMITS
While setting the High/Low alarm limits, if no alarm
limit is desired for the value being displayed, press
the MIN/MAX button, “--.-” will appear on the display
indicating the alarm for the value is disabled.
Once a high and/or low alarm value has been set for
a reading, the alarm will be enabled. appearing
to the left of the display value indicates that an alarm
is enabled for that reading. Each readings High/Low
alarm limit may be independently enabled/disabled
allowing the selection of high alarm only, low alarm
only, or both high and low alarm monitoring.
appearing will indicate that either a high alarm, low
alarm or both high and low alarms are enabled for
that reading.
1. Press and hold the SET button for 3 seconds,
release the SET button when “SET” appears at
the bottom right side of the display in place of the
time-of-day.
2. The low alarm limit for the ambient temperature
will appear. The alarm value and “Lo” will ash
on the display.
3. Press the or button to increment the
display to the desired low alarm value.
4. Press the SET button. The high alarm limit for
the ambient temperature will appear. The alarm
value and “Hi” will ash on the display.
5. Press the or button to increment the
display to the desired high alarm value.
6. Press the SET button. The low alarm limit for the
ambient relative humidity will appear. The alarm
value and “Lo” will ash on the display.
7. Press the or button to increment the
display to the desired low alarm value.
8. Press the SET button. The high alarm limit for
the ambient relative humidity will appear. The
alarm value and “Hi” will ash on the display.
9. Press the or button to increment the
display to the desired high alarm value.
10. Press the SET button. The low alarm limit for the
dew point temperature will appear. The alarm
value and “Lo” will ash on the display.
11. Press the or button to increment the
display to the desired low alarm value.
12. Press the SET button. The high alarm limit for
the dew point temperature will appear. The alarm
value and “Hi” will be ashing on the display.
13. Press the or button to increment the
display to the desired high alarm value.
14. Press the SET button. The data logging rate will
ash on the display (see the “Adjusting Data
Logger Recording Rate” section).
15. Press the SET button to return to the current
temperature/humidity display.
VIEWING HIGH/LOW ALARM LIMITS
Once alarm limits have been set, follow the steps in
the “Setting High/Low Alarm Limits” section to view
the values that have been set. If no changes are
desired, do not press the or buttons, simply
press the SET button to move to the next value.
DISABLE/ENABLE HIGH/LOW ALARM LIMITS
Once alarm limits have been set, follow the steps in
the “Setting High/Low Alarm Limits” section to view
the values that have been set. Press the SET button
to cycle through the values that have been set.
An alarm limit may be disabled while being displayed
by pressing the MIN/MAX button, “- -.-” will appear
on the display indicating the alarm for the value is
disabled. When an alarm value is disabled, the
alarm value previously set will be cleared.
To enable an alarm that has been disabled, while
viewing the disabled alarm (“- -.-”), press the or
button to increment the value to the desired alarm
value. With an alarm value selected (“- -.-” no longer
appearing), the alarm will be enabled.
Each High/Low alarm limit may be independently
enabled/disabled allowing selecting of high alarm
only, low alarm only, or both high and low alarm
monitoring.
ALARM SOUNDING
With an alarm enabled, when a reading being
measured is outside the alarm limit that has been
set (equal to or lower than the low alarm set point, or
equal to or greater than the high alarm set point):
will ash on the display next to the
measurement that is alarming (ambient
temperature, ambient relative humidity or probe
temperature). will continue to ash next to the
measurement until the measured value returns to
an in-range condition.
The red LED will ash continuously for the rst 10
seconds, and will then ash every 10 seconds
until the measured value returns to an in-range
condition.
The audible alarm will sound continuously for the
rst 10 seconds, and will then sound every 10
seconds until the measured value returns to an
in-range condition.
While alarming, the sound may be turned off by
pressing the SET button (“Off” will appear on the
display), it is not recommended to turn off the
alarm sound. This will turn off the alarm sound
and no audible alarms will sound while the alarm
sound is turned off. The alarm sound may only be
turned back on by reaching and alarm condition and
pressing the SET button to turn the sound back on;
therefore, it is not recommended to turn off the
alarm sound. The ashing and red LED alarm
are not affected by this setting.
DATA LOGGER
The built-in data logger is capable of utilizing SD
(2GB maximum) and SDHC (4GB maximum)
memory cards.
The sampling rate for the data logger may be set to
record readings from every 1 minute to every 720
minutes in 1 minute increments. A 1 minute interval
will record 1440 readings/day. A 720 minute interval
will record 2 readings/day.
SD CARD SELECTION
The supplied 256MB SD memory card is ready for
immediate use. When utilizing 3rd party SD/SDHC
cards, insure that they are formatted as FAT16 or
FAT32, NTFS format is not supported. SD cards up
to 2GB and SDHC cards up to 4GB are supported.
Cards over 4GB and MMC/SDXC cards are not
supported.
The supplied 256MB SD card is capable of storing
5.9 million readings (over 10 years of readings at 1
minute intervals).
Higher capacity memory cards require more power
to operate; therefore, larger cards will exhaust the
batteries more rapidly (see the “AC Power Supply/
Battery Power” section).
INSERTING/REMOVING SD MEMORY CARD
SD card door
Lift up here to open SD card door
Proper SD memory card orientation for inserting
To remove the SD memory card:
1. Turn off data logging (see the “Turning On/Off
Data Logging” section).
2. Turn the meter off, by pressing the ON/OFF button.
3. Open the SD card door by lifting up on the bottom
of the door.
4. Push the SD memory card until it “clicks”, then
release to allow the card to be ejected.
To insert the SD memory card:
1. Turn the meter off, by pressing the ON/OFF
button.
2. Open the SD card door by lifting up on the bottom
of the door.
3. Make certain that the SD memory card is oriented
properly (see image) and insert by pushing the
card until it “clicks”.
4. Close the SD card door by snapping it shut.
5. Turn the meter on by pressing the ON/OFF button.
6. Turn data logging on, if desired (see the “Turning
On/Off Data Logging” section).