Traceable digital oxygen meter instructions, Control company – Control Company 4320 DIGITAL OXYGEN METER User Manual
Page 2
©2007
Control Company. 92-4320-00 Rev. 0 103007
RS—232 PC SERIAL INTERFACE
This unit features RS—232 output. A RS—232 OUTPUT
SOCKET (4-10, fi g. 1) is located on the top of the unit. To
utilize this feature, connect the unit to a PC with the Data
Acquisition Accessory. (SEE ACCESSORIES SECTION).
Data is displayed in a 16 digit stream: D15 through D0.
Each digit indicates the following:
D0
end word
D1 to D4
Upper Display Reading, D1=LSD, D4=MSD
D5 to D8
Lower Display Reading, D5=LSD, D8=MSD
D9
Decimal Point (DP) for Upper Display:
0= no DP, 1=1 DP, 2= 2 DP , 3= 3 DP
D10
Decimal Point (DP) for Lower Display:
0= no DP, 1= 1 DP, 2=2 DP ,3= 3DP
D11/D12
Annunciator for Upper Display:
00= No Symbol, 01= °C, 02= °F,
03= %, 04= %RH, 05= % PH, 06= %O
2
07= mg/L, 08= m/s, 09= knots, 10= Km/h,
11= Ft/mm, 12= mph, 13= ųS, 14=mS
15= Lux, 16= Ft-cd, 17=dB, 18=mV
D13
Annunciator for lower display:
0= No symbol, 1= °C, 2= °F
D14
Reading Polarity for the display:
0= both upper and lower displays are “+”,
1= upper is “–” and lower is “+”
2= upper is “+” and lower is “–”,
3= both upper and lower displays are “–”.
D15
start word
BATTERY LIFE
If the letters “LBT” appear on the left corner of the
display, it indicates the battery is low and needs to be
replaced. To replace the battery, slide the battery cover
located on the back of the unit away from the unit.
Remove the old battery and replace it with a new 9-Volt
alkaline battery. Use an alkaline battery, NOT a regular or
heavy duty battery. Properly connect the battery. Replace
the battery cover. Incorrectly installed batteries may
damage electronics.
ALL OPERATIONAL DIFFICULTIES
If this oxygen meter does not function properly for any
reason, please replace the battery with a new 9-Volt
alkaline batteries (see Low Battery section). Low battery
power can occasionally cause any number of “apparent”
operational diffi culties. Replacing the batteries with new
fresh batteries will solve most diffi culties.
ACCESSORIES
Cat. No. 4138
Easy-Use™ Accessory Adaptor 115 VAC
Bonus Free Disk This demonstration disk shows the
data acquisition capabilities of this digital product when
connected to a computer. See Cat No. 4136 below to
order complete Data Acquisition System.
Cat No, 4136
Data Acquisition System Accessory Complete
DAS—3™ Data Acquisition System captures, displays,
and stores readings on any IBM-compatible PC.
Information can be imported into databases. Includes 3½
inch diskette and 5’ serial cable with D9F plug.
Cat No. 4325
Data Logger Complete DAS-4™ System captures and
stores up to 8000 bytes (over 1000 readings) from all
meters. Reading may be taken at intervals from 1 second
to 99 hours. Stored readings may be downloaded to
any PC and viewed. Can be read “as is” or imported
to spreadsheets, databases, and statistical programs.
Supplied with 36-inch serial cable with D9F computer
plug, 3½ inch diskette (Windows
®
), and four AA alkaline
batteries. Size is 5 x 3 x 1 inches. Weight is 7 ounces.
Cat. No. 4326
Accessory Adaptor 115 VAC for Data Logger
WARRANTY, SERVICE, OR RECALIBRATION
For warranty, service, or recalibration, contact:
CONTROL COMPANY
4455 Rex Road
Friendswood, Texas 77546 USA
Ph. 281 482-1714
Fax 281 482-9448
E-mail [email protected]
www.control3.com
Control Company is ISO 9001
Quality-Certified by DNV and
ISO 17025 accredited as a
Calibration Laboratory by A2LA.