8.2 Power Settings
The power limit settings (parameter
OP.Hi
) for the STF models on various supply voltages are as
follows. The figures represent the maximum percentage of time that controlled power is supplied
to the elements. Do not attempt to “improve performance” by setting a value higher than the
correct one from the table. TZF has different power settings for the three zones.
* 5000W model with 4 elements up to late 2000.
** model with 4 elements up to 1995.
STF -/610 were known as CTF before late 1996.
Model
Volts:
110V
115V
120V
200V
208V
220V
230V
240V
380V
400V
415V
49
45
41
60
55
49
45
41
78
72
64
59
54
64
59
54
100
92
82
75
69
59
54
49
74
68
78
71
65
78
71
65
84
77
70
61|78|61 57|72|57 51|65|51 46|59|46 43|54|43
46
43
38
35
32
62
57
51
47
43
51
47
43
76
70
63
57
53
63
57
53
59|45|59 54|41|54 50|37|50
Model
Volts:
200V
208V
220V
230V
240V
380V
400V
415V
42
39
34
31
29
34
31
29
61|78|61 57|72|57 51|65|51 46|59|46 43|54|43 51|65|51 46|59|46 43|54|43
37
34
30
28
25
30
28
25
34
31
28
29
27
28
29
27
74|75|74 69|69|69 61|62|61 56|56|56 52|52|52 61|62|61 56|56|56 52|52|52
TZF 16/610
STF 16/610
2-phase+N
3-phase+N
Single phase
STF 16/610
STF 15/610
TZF 15/610
STF 16/450
CTF 15/--/610 **
STF 15/180
STF 15/450
STF 15/--/450 *
STF 15/610
3-phase delta
TZF 15/610
STF 16/180
STF 16/450
TZF 16/610