5.7 Fuse Replacement
Fuses are marked on the circuit diagram (section 7.0) with type codes, e.g. F1, F2. A list of the
correct fuses is given in section 8.1. Depending on model and voltage, the different fuse types may
or may not be fitted.
If any fuse has failed, it is advisable for an electrician to check the internal circuits. Replace any
failed fuses with the correct type. For safety reasons do not fit larger capacity fuses without first
consulting Carbolite.
The fuses are near the cable entry point, and access is by removal of the back panel.