Dual Cool DCX-700 Instructions 9-30-13
1.3
Comb Specifications for Hand Cast Gels
1.4
Safety
Power to the Dual Cool Electrophoresis System is to be supplied by an external DC
voltage power supply that must be ground isolated so that the DC voltage output
floats with respect to ground. For any power supply used, the maximum specified
operating parameters for the units are:
Maximum Limits
300 VDC
30 watts power
190mA current
60°C ambient temperature
Current to the unit, provided from the external power supply, must enter the unit
through the lid assembly, providing a safety interlock to the user. Current to the unit
is broken when the lid is removed.
Do not attempt to use the unit without the
safety lid, and always turn the power supply off before removing the lid, or when
working with the unit in any way. Follow safety precautions specified by the
power supply manufacuturer.
COMB OPTIONS, material: PTFE, tooth depth: 9.5mm, overall length: 7.9cm
Cat. #
# of
Teeth
Thickness of
Teeth (mm)
Width of
Teeth (mm)
Recommended Max.
Sample Vol./Well (µl)
*
MVX-0701
1
0.75
66.6
356
MVX-0710
10
0.75
5.23
28
MVX-0714
14
0.75
3.3
19
MVX-0716
16
0.75
2.92
16
MVX-1001
1
1.0
66.6
474
MVX-1010
10
1.0
5.23
37
MVX-1014
14
1.0
3.3
24
MVX-1016
16
1.0
2.92
21
MVX-1501
1
1.5
66.6
712
MVX-1510
10
1.5
5.23
56
MVX-1514
14
1.5
3.3
35
MVX-1516
16
1.5
2.92
31
*
Maximum loading volume is calculated by taking 75% of the total tooth volume.