Dual Cool DCX-700 Instructions 9-30-13

1.3

Comb Specifications for Hand Cast Gels

1.4

Safety

Power to the Dual Cool Electrophoresis System is to be supplied by an external DC
voltage power supply that must be ground isolated so that the DC voltage output
floats with respect to ground. For any power supply used, the maximum specified
operating parameters for the units are:

Maximum Limits

300 VDC

30 watts power

190mA current

60°C ambient temperature


Current to the unit, provided from the external power supply, must enter the unit
through the lid assembly, providing a safety interlock to the user. Current to the unit
is broken when the lid is removed.

Do not attempt to use the unit without the

safety lid, and always turn the power supply off before removing the lid, or when
working with the unit in any way. Follow safety precautions specified by the
power supply manufacuturer.

COMB OPTIONS, material: PTFE, tooth depth: 9.5mm, overall length: 7.9cm

Cat. #

# of

Teeth

Thickness of

Teeth (mm)

Width of

Teeth (mm)

Recommended Max.

Sample Vol./Well (µl)

*

MVX-0701

1

0.75

66.6

356

MVX-0710

10

0.75

5.23

28

MVX-0714

14

0.75

3.3

19

MVX-0716

16

0.75

2.92

16

MVX-1001

1

1.0

66.6

474

MVX-1010

10

1.0

5.23

37

MVX-1014

14

1.0

3.3

24

MVX-1016

16

1.0

2.92

21

MVX-1501

1

1.5

66.6

712

MVX-1510

10

1.5

5.23

56

MVX-1514

14

1.5

3.3

35

MVX-1516

16

1.5

2.92

31

*

Maximum loading volume is calculated by taking 75% of the total tooth volume.