1 introduction and features, 2 specifications – C.B.S. Scientific DCX-700 User Manual
Page 5
5
www.cbsscientific.com
SECTION 1
General Information
1.1
Introduction and Features
C.B.S. Scientific offers the Dual Cool Electrophoresis System for performing SDS-Page,
acrylamide-nucleic acid separations and electro-blotting. The dual unit provides the
capability of running or blotting two gels simultaneously under identical temperature
controlled buffer conditions. Units include 2 freezer blocks, 2 blotting cassettes, 2 foam
pads, and a white plastic reservoir conversion plate to allow for single runs.
This system is designed primarily to be used with a variety of precast gels. (See Table
1). Hand-cast gels can also be run in this unit with the use of accessory combs, plates
and spacers.
Table 1: Features of Vertical Mini-Gel Systems
1.2
Specifications
Constructions:
Buffer chamber, safety cover, core
Polycarbonate/Polysulphone
Electrodes
Pure Platinum wire .010” diameter
Power cords
Silicone
rated
7500VDC,
200mA,
65°C
Combs
Teflon
Glass plates
Soda-lime float glass & Borosilicate
Spacers
PVC
Clamps
Polypropylene,
stainless
steel
Safety Certification
EN61010-1-1993 (IEC1010-1)
Table 2: Specifications
Model #
DCX-700
Shipping Weight
6 lbs
Overall Size W x D x H (cm)
16.5 x 15.2 x 21.6
Recommended buffer volumes:
Cathode reservoir
Anode reservoir
Total buffer volume for 2 blots
200mls
400mls (minimum) / 800mls (maximum)
940mls
Slab gel run time
30-90 minutes
Blot run time
75 minutes
Handcast glass plate size
10cm x 10cm
Voltage limit
300 V
Model #
Single or
Dual
Plate Dimensions
(w x h)
Compatible
Precast Gels
Cooling
DCX-700
Dual
10cm x 10cm
ClearPAGE™
Expedeon™
Invitrogen™
1 or 2 freezer blocks
with stirring
10cm x 8cm
ClearPAGE™
Expedeon™
10cm x 9cm (requires
adapter #DCX-AP-109)
Lonza™