3 references, Electro-blotting – C.B.S. Scientific DCX-700 User Manual
Page 18
18
Dual Cool DCX-700 Instructions 9-30-13
6.2
Electro-Blotting
6.3 References
1.
Hames, B.D. (ed.) (1998). Gel Electrophoresis of Proteins. A Practical
Approach. 3
rd
edn. Oxford University Press, Oxford. Ch.
1,3.
2.
Sambrook, J., Fritsch, Russell, D. (2001). Molecular Cloning. A Laboratory
Manual. 3
rd
edn. Cold Spring Harbor Laboratory Press, Cold Spring Harbor,
New
York. A8.40-A8.55
3.
Ausubel, F.M., Brent, R., Kingston, R.E., Moore, D.D., Seidman, J.G., Smith,
J.A.,
Struhl, K. (ed) (1993). Current Protocols in Molecular Biology. Vol. 2,
Greene Publishing Associates, Inc. and John Wiley & Sons, Inc., Ch.
10.
As a general recommendation, equilibrate gels (after running) with the diluted transfer
buffer for 5 to 10 minutes before transfer.
Blotting Buffer
ClearPAGE™ Transfer Buffer 10X/20X
cat. # FB82500
(see Section 6.4)
100ml
(1:10 dilution)
Methanol
200ml
Ultrapure Water
720ml
Typical Blotting Conditions for DCX-700
Power Supply Setting
200V constant
Blot Time
1.5 - 2.0 hours with stirring, cooling blocks
Expected Current
180mA / 1 gel
220mA / 2 gels