background image

Change adjustment

Eppendorf Xplorer

®

/Xplorer

®

plus

English (EN)

9

2.2.1

Reset to the factory setting

1. Move the selection dial to

Opt.

2. Select the

Adjustment

option with the rocker.

3. Press the

Select

softkey to open the adjustment.

4. Select the

Factory settings

option with the rocker.

5. Press the

Select

softkey to reset the adjustment to the factory setting.

6. Press the

Yes

softkey to confirm the confirmation request.

The factory setting will become active after a few seconds.

7. Turn the selection dial to any operating mode to exit the options.

The wrench symbol and the adjacent adjustment symbol have disappeared from the header.

2.3

Liquid type ethanol 75%

The factory setting is changed by a factor in order to allow 75% ethanol to be dispensed with higher
accuracy (smaller systematic error) with the Xplorer pipette. The internally used factor takes into account
the density. Perform pipetting with this adjustment in the

Pip operating mode as follows:

1. Hold the Xplorer pipette vertically and prewet the tip with liquid three times at room temperature. Make

sure that no outer wetting remains on the tip.

2. Use speed level 5 (standard setting) for aspiration and dispensing.

3. Dispense liquid at a slight angle against the wall of the tube of the destination tube.

4. About 2 seconds after dispensing keep the rocker pressed to trigger the blow-out (blow). Only release

the rocker after leaving the dispensing vessel.

5. Eject the pipette tip.

2.4

Liquid type glycerol 50%

The factory setting is changed by a factor in order to allow 50

% glycerol to be dispensed with higher

accuracy (smaller systematic error). The internally used factor takes into account the density of the 50%
glycerol and the same work technique as described for ethanol

75% in the

Pip mode.

Before routine use, check whether this adjustment meets your requirements. If the
concentration is different, check whether the 1

3

point

adjustment would be a better

solution.

Before routine use, check whether this adjustment meets your requirements. If the
concentration is different, check whether the 1

3

point

adjustment would be the better

solution.