Eppendorf Xplorer

®

/Xplorer

®

plus

English (EN)

21

3.2.2

Xplorer pipette, multi-channel

pipette
Increment

Pipette tip
Color code
Volume range
Length

Testing
volume

Error limits

Error

Systematic error

Random error

µL

µL

±%

±µL

±%

±µL

0.5

μL – 10

μL

Increment:
0.01

μL

Medium gray
0.1

μL – 20

μL

40 mm

1

μL

±5.0

±0.05

±3.0

±0.03

5

μL

±3.0

±0.15

±1.5

±0.075

10 μL

±2.0

±0.2

±0.8

±0.08

5

μL – 100

μL

Increment:
0.1

μL

Yellow
2

μL

–200

μL

53

mm

10 μL

±2.0

±0.2

±2.0

±0.2

50

μL

±1.0

±0.5

±0.8

±0.4

100

μL

±0.8

±0.8

±0.25

±0.25

15

μL – 300

μL

Increment:
0.2

μL

Orange
15

μL

–300

μL

55

mm

30

μL

±2.5

±0.75

±1.0

±0.3

150

μL

±1.0

±1.5

±0.5

±0.75

300

μL

±0.6

±1.8

±0.25

±0.75

50

μL – 1200

μL

Increment:
1

μL

Green
50

μL–1250

μL

76

mm

120

μL

±6.0

±7.2

±0.9

±1.08

600

μL

±2.7

±16.2

±0.4

±2.4

1200

μL

±1.2

±14.2

±0.3

±3.6